MODEL EDM365
RUGGED OUTDOOR PM MONITOR
The Stand-Alone Brother of the Approved EDM180
The mobile environmental dust monitor EDM 365 was developed and designed by Grimm for the mobile industrial use. This high-tech stand-alone system EDM 365 satisfies customers world-wide with its high precision and the low maintenance and operating costs. The EDM 365 detects airborne particles in real-time without the loss of the semi-volatile compounds.
Every single aerosol particle is detected in the optical measurement cell. This precise and reliable single particle count allows for a simultaneous measurement of the dust mass fractions PM10, PM2.5, PM1 and also the particle size distribution in 31 size channels.
The outdoor weather protection housing is thermo-isolated with an automatic heating and proportional ventilation.
Overview
- equivalent to norms US-EPA, UK-MCERTS, CN-CMA
- weather-protected system in powder-coated stainless steel housing
- continuous use in harshest environments
- real-time measurement of PM10, PM2.5, PM1, total counts (TC), and particle number distribution
- fully automated monitoring system with remote access
- extremely energy-efficient, low maintenance, no consumables
- no loss of semi-volatile compounds
- no radioactive source, insensitive to vibrations
- versatile data acquisition and communication (GSM data logger)
- self-test of all optical and pneumatic components for high quality standards
- rinsing air for protecting laser and detector in optical cell
- temperature and relative humidity sensors
- total inlet flow analyzed in optical cell
- excellent counting statistics and reproducibility at low and high dust concentrations
Applications
- Continuous PM measurements where no container can be placed
- AMS for PM networks
- Industrial surveillance
- epidemiological studies
- monitoring of construction and mining sites
Technical Data
measured mass fractions
|
PM10, PM2.5, PM1
|
particle size range
|
0.25 - 32 μm
|
size channels
|
31
|
particle number
|
0 - 3 000 000 particles/liter
|
reproducibility
|
>97% of total measuring range
|
detection principle
|
light scattering at single particles with diode laser; detection volume aerodynamically focused, no border zone error
|
optical cell
|
diode laser 660 nm
|
detector | fast signal processing, 2 x 16 raw data channels |
time resolution
|
selectable storage intervals 6 s; 1, 5, 10, 15, 30, 60 min
|
sample flow rate
|
1.2 L/min, ± 3% constant due to self-regulation
|
rinsing air
|
0.4 L/min, protection of laser optics, reference air for self-test
|
sampling inlet | isothermal humidity extraction via Nafion membrane, sensor-controlled, without loss of semi-volatile compounds (SVC) |
operation
|
keypad, data logger or PC with GRIMM software or HyperTerminal
|
interfaces
|
Ethernet & GSM via data logger, RS-232 (GESYTEC)
|
analog input
|
1 port (0 - 10 V) for auxiliary sensors
|
digital input | 1 port for sensors 157L, 158L, 159L (see accessories) |
power supply
|
in: 230 V/50 Hz; optional 115 V/60 Hz
|
power consumption
|
Pmax = 150 W, Imax: 1.4 A
|
temperature range | -20 to +40°C (-4 to 104°F), non-condensing |
absolute pressure range | 900 - 1100 mbar |
dimensions
| 48.9 x 70 x 27.1 cm (19.3 x 27.6 x 10.7 in) without sample pipe |
weight
|
36 kg (79.4 lbs)
|
Accessories
1178
|
GRIMM software
|
1146
|
GPS sensor
|
157L | sensor for temperature, relative humidity, and barometric pressure |
158L | sensor for temperature, relative humidity, barometric pressure, wind speed, and wind velocity |
159L | sensor for temperature, relative humidity, barometric pressure, wind speed, wind velocity, and precipitation |
365-DL | Data logger for our measurement equipment including the capability to transmit all data live via GSM onto a password-protected platform |
SOFTWARE
GRIMM developed a powerful software to visualize, process, and import as well as export your valuable measurement data.
This easy-to-use program is compatible to Windows XP and all later versions.
Please contact us for your for free trial version if you already own a GRIMM measurement device or more information if you are in the process of finding your measurement solution.