MODEL EDM164
A GENUINE NEAR-REFERENCE METHOD
Small & Compact PM Monitor for Continuous Outdoor Monitoring
The EDM 164, in its compact and mobile weather housing, performs reliable optical detection for counting and classifying dust particles. The EDM 164 is applicable for short and long-term continuous monitoring of dust pollution and enables a real-time data analysis of dust and meteorological measurement data.
This configuration places the EDM 164 in the leading position of the mobile environmental dust monitoring for such applications as dust mass, PM values, total counts and particle size distribution. The EDM 164 is the optimal solution for reliable environmental monitoring, e.g. PM monitoring, source identification, or pollution control at construction and mining sites.
Overview
- fully automatic monitoring system with remote access
- extremely energy-efficient, low maintenance
- no consumables
- real-time monitoring of PM values (PM10, PM2.5, PM1) and particle number
- additional information on particle size and mass distribution
- high precision over 31 size channels
- versatile data aquisition and communication (data logger with GSM via internet)
- self-test of all optical and pneumatic components for high quality standards
- rinsing air for protecting laser and detector in optical cell
- meteorological sensors
- 11 size channels < 1 μm for precise submicron detection
- total inlet flow volume (1.2 L/min) entirely analyzed in optical cell
- excellent counting statistics and reproducibility at low and high dust concentrations
Applications
- mobile and simultaneous monitoring of PM10, PM2.5, and PM1
- hot spot monitoring
- public site and urban monitoring
- source identification
- quantification of diffusive emissions according to VDI 4285 part 3
- early warning system for forest fires
Technical Data
measured mass fractions | PM10, PM2.5, PM1, TC |
particle size range | 0.25 - 32 μm |
size channels | 31 |
particle number | 0 - 3 000 000 particles/liter |
dust mass | 0 μg/m3 - 100 000 µg/m3 |
reproducibility | >97% of total measuring range |
optionally | GPS positioning, TSP (Total Suspended Particles), TC (Total Counts), and particle number for all size channels (size distribution) |
detection principle | light scattering at single particles with diode laser; detection volume aerodynamically focused, no border zone error |
optical cell | diode laser 660 nm |
detector | fast signal processing, 2 x 16 raw data channels |
time resolution | selectable storage intervals 6 s; 1, 5, 10, 15, 30, 60 min |
sample flow rate | 1.2 L/min, ± 3% constant due to self-regulation |
rinsing air | 0.4 L/min, protection of laser optics, reference air for self-test |
sampling inlet | heated, constant above ambient temperature |
operation
|
keypad, data logger or PC with GRIMM software (wireless or data
|
interfaces
|
data logger (4 x RS-232, RS-485, Ethernet, μSD, GSM) or RS-232
|
analog input
|
1 port (0 - 10 V) for auxiliary sensors
|
power supply
|
in: 110 – 230 VAC, 50 – 60 Hz
|
power consumption
|
22 W standard, Imax: 1.4 A
|
temperature range | - 20 to +60°C (-4 – 140°F), RH < 95%, non-condensing |
absolute pressure range | 900 - 1100 mbar |
dimensions
| housing only: 21 x 45 x 44 cm (8.3 x 17.7 x 17.3 in); with meteorological sensor 157L / 158L / 159L 23 x 51 x 73 / 79 / 85 cm (9.1 x 20.0 x 28.7 / 31.1 / 33.5 in) |
weight
|
20 kg (44.1 lbs)
|
Accessories
1178
|
advanced GRIMM software for EDM164
|
1146
|
GPS sensor
|
158-TRH Kit
| temperature and relative humidity sensor, with 3.5 meter cable and rain protection housing |
157L | sensor for temperature, relative humidity, and barometric pressure |
158L | sensor for temperature, relative humidity, barometric pressure, wind speed, and wind velocity |
159L | sensor for temperature, relative humidity, barometric pressure, wind speed, wind velocity, and precipitation |
160DL | Data logger for our measurement equipment including the capability to transmit all data live via GSM onto a password-protected platform |
SOFTWARE
GRIMM developed a powerful software to visualize, process, and import as well as export your valuable measurement data.
This easy-to-use program is compatible to Windows XP and all later versions.
Please contact us for your for free trial version if you already own a GRIMM measurement device or more information if you are in the process of finding your measurement solution.