Design and performance of an optimized electrical diffusion battery



Back to list of publications

Title:Design and performance of an optimized electrical diffusion battery
Author:
Date of Publication:2008
Hyperlink:www.elsevier.com/locate/jaerosci
Product:5601
Abstract:

Standard aerosol instruments to measure particle size distributions in the ultrafine size range are large and heavy. We built a small electrical  diffusion battery, which can be carried in a backpack and thusmakes size-resolved short-term personal aerosol exposition measurements possible. The  instrument was designed for maximal measurable particle size range and long maintenance intervals. The optimum number of stages for the diffusion battery was found with a Monte-Carlo simulation. To assess the instrument's performance, we compared size distributions measured with the electrical diffusion battery to those obtained with standard aerosol instruments (the scanning mobility particle sizer and the fast mobility particle sizer of TSI Inc.). In general, the readings of the electrical diffusion battery and those of standard instruments agree well (to within 10–20%).